Appendix b technical specifications, B.1 battery physical b.2 battery environmental – ZOLL AutoPulse Rev C User Manual

Page 29

Advertising
background image

P/N. 10762-001 Rev. 8

Page B-1

Battery Charger User Guide

Appendix B Technical Specifications

The specifications provided in this appendix apply to the AutoPulse Power System.

B.1

Battery Physical

B.2

Battery Environmental

Table B-1

Battery Specifications

Category

Specifications

Size (L×W×H)

11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm).

Weight

5.1 lbs. (2.3 kg).

Type

Rechargeable Nickel-Metal Hydride (NiMH)

Battery voltage (nominal)

32.4V

Capacity

3200 mAh (typical)

Initial Battery runtime
(nominal patient)

30 minutes (typical)

Maximum Battery charge
time

Less than 4

1

/

4

hours at 77ºF (25ºC)

Battery test-cycle time

Less than 10 hours per test-cycle session; up to three consecutive sessions
possible.

Required replacement interval 100 full charge/discharge cycles.

Note:

The Battery will not operate after 100 full charge/discharge cycles.

Table B-2

Battery Specifications (Page 1 of 2)

Category

Specifications

Operating temperature

+32º to +113ºF (0º to +45ºC) ambient installed in device

Charge temperature

+41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)

Storage temperature

-4º to +77ºF (-20º to +25ºC) ambient for less than six months (may require
test-cycle to meet performance characteristics)

+77º to +95ºF (+25º to +35ºC) ambient for less than two months (may
require test-cycle to meet performance characteristics)

Operating altitude

0 to 15,000 ft. (0 to 4,572 m)

Enclosure protection

Meets IP22 per IEC 60529

Shock

Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)

Advertising