P/N. 10762-001 Rev. 8

Battery Charger User Guide

B.3

Battery Charger Physical

B.4

Battery Charger Environmental

Vibration

Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s

2

)

Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)

Free fall

Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.

Table B-3

Battery EMI/EMC Specifications

Title

Standard

Electrostatic discharge

IEC 61000-4-2, Level 3

Radiated emissions

CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A

Table B-4

Battery Charger Physical Specifications

Category

Specifications

Size (L×W×H)

15 in. by 9.75 in. by 9.1 in. (38 cm by 25 cm by 23 cm).

Weight

10 lbs. (4.5 kg)

Operating input voltage

100 to 240V AC

Operating input frequency

50/60 Hz

Input current

2.0 Amps (maximum)

Maximum Battery charge
time

Less than 4

1

/

4

hours (at 77ºF [25ºC])

Fuses

User-replaceable, T2.0A 250V AC (2 required)

Table B-5

Battery Charger Environmental Specifications (Page 1 of 2)

Category

Specifications

Operating temperature

+41º to +95ºF (5º to +35ºC) (68º to 77ºF [20º to 25ºC] preferred)

Storage temperature

-40º to +158ºF (-40º to +70ºC)

Relative humidity

5% to 95%, non-condensing.

Operating altitude

0 to 10,000 ft. (0 to 3,048 m)

Enclosure protection

Meets IP22 per IEC 60529