Table a-2 pro – ZOLL AutoPulse Rev C User Manual

Page 28

Advertising
background image

Page A-2

P/N. 10762-001 Rev. 8

Battery Charger User Guide

Table A-2

Battery Charger Troubleshooting Procedures

Symptom

Possible Cause

Recommended Action

Battery Charger’s green
POWER LED is not
illuminated.

Battery Charger’s
alternating current (AC)
power cord is not plugged
in.

Refer to Chapter 2, “Setting Up the AutoPulse Battery
Charger” for more information.

Battery Charger’s green
POWER LED is not
illuminated.

Blown fuse.

Refer to Section 5.3.2, “Replacing a Battery Charger
Fuse” for more information.

Charging a Battery takes
much longer than 4

1

/

4

hours.

The ambient temperature
around the Battery Charger
is too warm.

Make sure that the Battery Charger is located in
an environment where temperatures do not reach
above 113˚F (45˚C).

Make sure that the Battery Charger’s vents are
not blocked.

Make sure that the Battery Charger has adequate
ventilation.

Battery Charger’s red
FAIL LED is illuminated
during the charging cycle.

The Battery has failed the
charge cycle.

Press TEST button to start the test cycle. The TEST
LED will illuminate and the Battery Charger will
attempt to restore the Battery.

Battery Charger’s red
FAIL LED is illuminated
during the test cycle.

The Battery has failed the
test cycle.

Remove the Battery from the Battery Charger.
Perform a Battery status check:

1.

If the Battery status LED is flashing red, the
Battery has failed. Replace the Battery. Refer to
Section 5.2.2, “Reaching the End of Battery
Service Life” for more information.

2.

If no status LEDs illuminate when you press the
Battery’s Status Check button, the Battery has
failed. Replace the Battery. Refer to Section
5.2.2, “Reaching the End of Battery Service Life”
for more information.

3.

If the Battery status LED is green or amber,
contact ZOLL Circulation.

Advertising