B.5 fcc statement – ZOLL AutoPulse Rev C User Manual

Page 31

Advertising
background image

P/N. 10762-001 Rev. 8

Page B-3

Battery Charger User Guide

B.5

FCC Statement

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.

B.6

Limited Warranty for AutoPulse Resuscitation System

ZOLL Circulation warrants to the initial Purchaser only that the “Warranted Product” purchased
hereunder will be free from defects in workmanship or materials, when given normal, proper, and
intended usage, for a specified period (“Warranty Period”) from the date of its initial shipment to
Purchaser. “Warranted Products” consist solely of those products whose description in this price list

Shock

Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)

Vibration

Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s

2

)

Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)

Free fall

Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.

Safety requirements

Safety certified to UL2601, CSA 22.2 No. 601.1-M90, EN60601-1

Table B-6

Battery Charger EMI/EMC Specifications

Title

Standard

Electrostatic discharge

IEC 61000-4-2, Level 3

RF electromagnetic fields
immunity

IEC 61000-4-3, Level 2

EFT/burst

IEC 61000-4-4, Level 2

Surge immunity

IEC 61000-4-5, Level 2

Conducted RF disturbances
immunity

IEC 61000-4-6, Class A

Dips, interruptions, and
variations

IEC 61000-4-11

Harmonics current emissions IEC 61000-3-2, Class A

Radiated emissions

CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A

Table B-5

Battery Charger Environmental Specifications (Page 2 of 2)

Category

Specifications

Advertising