ZOLL M Series Invasive Blood Pressure

7

The source of the displayed Heart Rate values depends upon the signals being monitored by the M
Series unit:

• If the ECG signal is present, the ECG heart rate displays.
• If the ECG signal is not present, but pulsatile IBP signal is present, the IBP pulse rate displays

(regardless of whether the SpO

2

signal is present). P1 takes precedence over P2.

• If neither the ECG nor pulsatile IBP signals are present, the SpO

2

pulse rate displays.

The IBP display area consists of:

• Symbols that represent the status of pressure measurements.
• Numbers that represent the measurement values (see Fig. 6 and Fig. 7):

Figure 6: IBP Data Display Area - ART, PA, P1

Figure 7: IBP Data Display Area - CVP, ICP, P2

Either channel (P1, P2) may be used for any type of pressure measurement (arterial, pulmonary arterial,
central venous, or intracranial). The M Series default unit of measure is millimeters of mercury
(mmHg), but you can configure the M Series to display kilopascals (kPa). See the M Series CCT
Configuration Guide
for instructions on how to configure alternate units of measure.

Channel Label

Alarm Status

Diastolic BP Value

Systolic BP Value

Mean BP Value

Unit of Measurement

Channel Label

Alarm Status

Mean Pressure Value

Unit of Measurement