ZOLL M Series Invasive Blood Pressure

Safety Summary (cont.)

• If the accuracy of measurements is suspect, first check the IBP waveform for artifact such as

catheter whip. Then check the catheter, tubing, transducer, cabling and M Series IBP option for
proper functioning.

• Do not immerse the M Series device, batteries, cables, or transducers in water, solvents, or cleaning

solutions.

• If the M Series device gets wet accidentally, wipe it down with a clean towel.
• Do not sterilize M Series unit or accessories except as specifically recommended in ZOLL

manuals.

• Connect the ECG-out jack and modem (if available) only to other equipment with galvanically

isolated circuits.

• The cable and IBP transducer(s) are defibrillator proof. Using the IBP option introduces no risk of

shock due to defibrillation. Using the IBP option introduces no risk of burns due to electro-surgery.

• Do not reuse single-use transducers.
• Check the transducer packaging to make certain the transducer is sterile.
• Check the expiration date on the transducer packaging. If it has expired, replace the transducer.
• If the device is damaged by age, physical or environmental abuse, check the unit using the

Maintenance Checklist to ensure its safety or remove the unit from use.

• Refer to the transducer manufacturer’s specifications for information regarding volume

displacement.

Indications for Use

The ZOLL M Series IBP option is indicated for measuring arterial, venous, and intracranial blood
pressures using invasive transducers with 5uV/V/mmHg sensitivity for resting patients in critical care
and transport. The M Series IBP option is intended for use with adult and pediatric patients only.

Contraindications for Use

The ZOLL M Series IBP option is not indicated for use on neonatal patients.