Setting alarms – ZOLL M Series CCT Defibrillator Rev D User Manual

Page 17

Advertising
background image

ZOLL M Series Invasive Blood Pressure

11

Setting Alarms

Before taking measurements, check that all IBP alarms settings are appropriate for the situation. You
can set the alarms for Pressure channel 1, Pressure channel 2 and NIBP separately. See M Series
Operator’s Guide
for instructions on setting alarms.

C

AUTION

!

Before the M Series with IBP option is initially put into service, the IBP alarm limit

defaults should be configured per institutional guidelines. See the M Series CCT
Configuration Guide
for further details.

The M Series IBP option includes blood pressure alarms for the following measurements:

Table 3: Alarm Limit Settings and Defaults - Central Venous Pressure (CVP)

Table 4: Alarm Limit Settings and Defaults - Intracranial Pressure (ICP)

Table 1: Alarm Limit Settings and Defaults - Arterial Pressure (ART)

Alarm Limit:

Default Setting:

Range (increments of 5):

High Systolic

160 mmHg; 21.3 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Systolic

90 mmHg; 12.0 kPa

-50 to 295 mmHg; -6.7 to 39.2 kPa

High Diastolic

110 mmHg; 14.6 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Diastolic

50 mmHg; 6.6 kPa

-50 to 295 mmHg;-6.7 to 39.2 kPa

High Mean

130 mmHg; 17.3 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Mean

60 mmHg; 8.0 kPa

-50 to 295 mmHg; -6.7 to 39.2 kPa

Table 2: Alarm Limit Settings and Defaults - Pulmonary Arterial Pressure (PA)

Alarm Limit:

Default Setting:

Range (increments of 5):

High Systolic

300 mmHg; 39.9 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Systolic

-50 mmHg; -6.7 kPa

-50 to 295 mmHg; -6.7 to 39.2 kPa

High Diastolic

300 mmHg; 39.9 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Diastolic

-50 mmHg; -6.7 kPa

-50 to 295 mmHg;-6.7 to 39.2 kPa

High Mean

300 mmHg; 39.9 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low Mean

-50 mmHg; -6.7 kPa

-50 to 295 mmHg; -6.7 to 39.2 kPa

Alarm Limit:

Default Setting:

Range (increments of 5):

High

300 mmHg; 39.9 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa

Low

-50 mmHg; -6.7 kPa

-50 to 295 mmHg; -6.7 to 39.2 kPa

Alarm Limit:

Default Setting:

Range (increments of 5):

High

300 mmHg; 39.9 kPa

-45 to 300 mmHg; -6.0 to 39.9 kPa