ZOLL M Series Invasive Blood Pressure

13

Printing IBP Data

See the M Series Operator’s Guide Recorder Operation section for instructions on how to print data. If
you are currently displaying IBP traces, press the RECORDER button to print a strip chart that includes
the following values for the currently displayed trace:

• One or two IBP traces (whichever is currently displayed).
• Systolic blood pressure.
• Diastolic blood pressure.
• Mean pressure.
• Time of measurement.

Maintaining the IBP Option

See the M Series Service Manual for instructions on maintaining the IBP option.

Performing the Daily Checkout Procedure

The following is a checkout procedure that should be performed daily to ensure that the IBP option is
functioning properly. In addition, this daily checkout procedure ensures that medical personnel
maintain familiarity with the proper use of the IBP option.
Use only 5uV/V/mmHg IBP simulators. See "Ordering Accessories" on page 16 for further details. For
more specific instructions on how to accomplish each step listed below, refer to the related section in
this manual.

To perform the daily checkout procedure:

1.

Connect the IBP simulator cable to the M Series unit and the IBP simulator.

2.

Set the simulator to zero.

3.

Zero the transducer.

4.

Set the simulator to simulate 50 mmHg.

5.

Verify that the M Series displays systolic, diastolic and mean pressures of 50 mmHg +/- 6

mmHg.

6.

Set the simulator to simulate 250 mmHg.

7.

Verify that the M Series displays systolic, diastolic and mean pressures of 250 mmHg +/- 6

mmHg.