ZOLL M Series Invasive Blood Pressure

3

4. Verify that neither the CHECK TRANSDUCER nor the REPLACE

TRANSDUCER messages display.
If desired, repeat steps 1 - 4 to connect a second transducer. The ZERO P1 or ZERO
P2 message displays. No pressure measurements will appear until the transducer
has been zeroed.

Setting up the transducer and catheter

Follow manufacturer’s directions for use and your hospital’s protocols on how to setup, prime and
flush the transducer and make connections to the pressure monitoring catheter.

Displaying the IBP Menu

Before you can zero the transducer, you must first display the IBP menu. You can display the IBP menu
without leaving Monitor, Defib or Pacer mode.
AEDs must be in manual mode to display the IBP menu. See the M Series Operator’s Guide for further
details on activating the manual override mode.

To display the IBP menu:

1.

Press the Param softkey. If the Param softkey label is not displayed, press the Return softkey

until “Param” displays.
The Parameter menu displays (see Fig. 2):

Figure 2: Parameter Menu

2.

Press the Select softkey until “IBP” highlights.

3.

Press the Enter softkey.

M Series displays the IBP menu (see Fig. 3):

Figure 3: IBP Menu

You can now zero the transducer(s).

NOTE

Unless otherwise noted, all the following procedures in this insert assume that you are already in the

IBP menu.