Specifications – ZOLL M Series CCT Defibrillator Rev D User Manual

Page 20

Advertising
background image

14

ZOLL M Series Invasive Blood Pressure

Specifications

For a list of alarm defaults and ranges, see "Setting Alarms" on page 11.

Table 5: General Specifications

Principle of Operation

Resistive Bridge transducer; 5uV/V/mmHg sensitivity
(transducers with 40uV/V/mmHg sensitivity are not
compatible)

Regulatory Standards

EN60601-2-34:1995

Table 6: IBP Measurement Specifications

Pressure Range

-50 to 300 mmHg after offset adjustment

Zero Offset Adjust Range

+/- 200 mmHg

Accuracy

Greater of +/- 2% measurement or +/- 2 mmHg, plus transducer
error

Baseline Stability

+/- 0.1 mmHg/

° C excluding transducer drift

Bandwidth

DC to 20 Hz

Noise

< 1 mmHg peak-to-peak

Transducer Impedances

Input: minimum 250

Ω Output: maximum 3000 Ω

Transducer Excitation

2.5 V DC

Table 7: Battery Operating Time Specifications (using ZOLL XL Battery Pack)

For a new, fully charged battery
pack at 20°C with IBP, EtCO

2

,

SpO

2,

NIBP and Temperature

options

60 defibrillator discharges at maximum energy (200J Biphasic),
or 1.75 hours minimum of continuous ECG, IBP (2 channels),
EtCO

2

, SpO

2

, and Temperature monitoring, or 2.00 hours of

continuous ECG monitoring/pacing at 60 mA, 70 beats/min.

Table 8: Environmental Specifications

Operating Temperature

0

°C to 50 °C

Storage and Shipping Temperature

-20

°C to 60 °C

Electromagnetic Immunity

AAMI DF-2: 1996, EN 61000-4-3: 1996, 15 V/m

Humidity

5% to 95% relative humidity, non-condensing

Operating Pressure

594 mBar to 1060 mBar