ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 137

Maintenance
7
Figure 8-1.
Location of Condensate Pan
.
5.
The pan may be washed in hot soapy water. Never use solvents or
abrasive cleaners on the pan. Dry the pan when finished and reinstall it
in the System.
IMPORTANT:
Do not operate the System without a condensate collection pan properly
installed. Failure to do so may allow water to accumulate on the floor under
the System.
Drain Coolant
It may become necessary to drain and clean the coolant well because the
coolant has become contaminated, or to prepare the System for shipment.
Required Tools and Materials
You will need the drain line and connector assembly (supplied with the System),
and a 2-liter waste container to perform this task.
Procedure
To drain the coolant well, follow these steps in the indicated order.
1.
Unplug the System.
600479-001 Rev 2 137/147