Operation
11
Figure 4-5. System Main Menu
Changing Target Temperature
1.
Press the Target Temp button once. If you are in RUN, you will be taken
to STANDBY. The Target Temperature value may only be changed in
STANDBY.
Figure 4-6. Change Target Temperature in RUN
2.
To change the target temperature, turn the knob until the desired
temperature is displayed. You may choose a temperature from 31º C to
38º C (87.8º F to 100.4º F). When the correct selection is displayed,
press the knob once to enter the selection.
600479-001 Rev 2 42/147