ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 80

Operation
49
Figure 4-60. Location of the Serial Interface Connector.
11. Start
the
TempTrend CSV program on the laptop computer. Refer to
Chapter 5 - TempTrend CSV.
12.
If a problem occurs during downloading, the System screen will display
the message “Download error. Please check external computer.” Check
the serial cable connections and the operation of the laptop computer
and choose “Try again.” If repeated failures occur, choose “Cancel.”
Repeated failures indicate a problem with the TempTrend program or the
System; contact your ZOLL representative for assistance.
Figure 4-61. Download Error Message.
13.
When the data download is complete, the System screen briefly displays
the message “Download Complete.” At this time, the patient trend data
saved from the improper shutdown is deleted.
600479-001 Rev 2 80/147