Installing the start-up kit – ZOLL Thermogard XP IVTM System Operation Manual User Manual
Page 63

Operation
32
Setup Sensor Checks
Air Trap
There is no Start-up Kit installed or there is a large amount of air in the
Start-up Kit chamber.
Roller Pump Lid
The clear plastic roller pump lid is not closed properly.
Check Prime Switch
The Prime Switch is being depressed (as when you use it to prime the
system) or has been jammed in the ‘on” position.
Coolant
The coolant level is low. Top up the coldwell with coolant.
Installing the Start-up Kit
Hang the Saline Bag
To install the System Start-up Kit, follow these steps in the indicated order.
1.
Obtain an IV bag or bottle of sterile normal saline solution. The bag or
bottle must contain no more than 0.5 liter (500 ml) of solution. Hang the
saline container on the hook mounted on the rear of the display post.
The container hangs inside the circumference of the handle.
Figure 4-33. 500ml Saline Bag on Hook.
The Top Cover
2.
Open the top cover of the System. Open the transparent top cover of the
roller pump.