D.

Differential Pressure Flowmeters With 4-20mA Output And With A
Linear Flow Relationship.
eg. D.P. transmitters with a square root extractor or VA meters.

......(8)

Q

VB

= S

VB

.

P

F

P

B

.

T

B

T

F

.

Z

B

Z

F

. A

Q

M

=

ρ

B

. Q

VB

Note that the pressure and temperature are still square rooted, even though
the flow signal A is not. This is because the output from the D. P.
transmitter is not truly volumetric, but will be affected by a change in
density of the gas being measured. Therefore, the equations relating to
differential pressure will apply.

E.

Dual Differential Pressure Flowmeters With 4-20mA Output.

To increase the range over which flow can be measured, two D. P.
transmitters with different spans can be connected across a common
orifice or other differential pressure device.

Equations 5 & 6 or 7 & 8 above would be used depending upon whether
the D. P. transmitters have square root extractors. Separate scaling using
these equations is then programmed for each transmitter.

At lower flowrates, transmitter 2 will be used as a basis of measurement
and at higher flowrates, transmitter 1 will be used. The crossover point
will occur when the input on transmitter 2 exceeds 20mA.