5.2 DEFINITIONS

GENERAL SETUP

Display Contrast

It is possible to adjust the contrast of the
display using the key to give optimum
viewability.

Total Units

Enables totals to be integrated at 1/1000 of the
flowrate by programming x 1000. For
example, if the flowrate is in scm/h the total
will be kscm/h (scm x 1000).

Timebase

Timebase selection will determine whether the

rate of flow is displayed in kg/day,
kg/hour...etc.

Front Access

If enabled, access to the Calibration routine is

possible via the front panel (via the Total and
Display keys). If disabled, access to the
Calibration routine is only possible by
connecting a link between terminals 1 and 2.

Front Reset

If disabled, the front panel reset key becomes
inoperable during normal operation.

GAS PARAMETERS

Default

Display

The flowrate and total which is normally

displayed (eg. Mass or Corrected Volume).

Calibration 59