2-2. cables, terminals, glands and conduit adapter, Figure 3-5b. how to install cable glands – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 16

Advertising
background image

8

IM 12D06D05-01E

High voltage section

Contact

(S1, S2)

output

cables

mA

cables

Contact

(S3, S4)

output

cables

Sensor

Cables

Input

contact

Power

cable

Suitable for cables with an outside diameter between 6 - 12 mm (0.24 - 0.47”)

Figure 3-5a. Cable glands diagram

3-2-2. Cables, Terminals, glands and conduit

adapter

ISC450G-A(D)-A

The ISC450 is supplied with terminals suitable

for the connection of finished wires in the size

range of 0.13 to 2.5 sq.mm. (26 to 14 AWG).

The cable glands supplied will form a tight seal

on cables with an outside diameter of 6 to 12

mm (0.24 to 0.47 inches). Unused cable entry

holes must be sealed with cable glands includ-

ing the blind plugs supplied.

ISC450G-A(D)-U

The ISC450 is supplied with terminals suitable

for the connection of finished wires in the size

range of 14- 26 AWG. The cable entry holes

are sealed with FM certified plugs. Prior to

cable entry the plugs can be removed with

allen key size 3/8” The cable conduit fittings

can be mounted in the holes of the housing as

required. The cable glands supplied with the

unit will give a tight seal on cables with outside

diameter of 0.24 to 0.47 inches.

Gland nut

O-ring

Gland

Close up plug

Contents:

6 X Gland M20

6 X Close up plug

6 X Gland nut M20

6 X O-ring 17.12 X 2.62 EPDM 70° sh.

NOTE: The glands must be installed properly

to meet IP66 and NEMA 4X rating.

Use close up plug for unused glands.

NOTE: Moisturize O-ring before assembling.

Figure 3-5b. How to install cable glands