Appendix 3, Calibration solutions for conductivity, Appendix 3, calibration solutions for conductivity – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 60

Advertising
background image

52

IM 12D06D05-01E

APPENDIX 3

Calibration solutions for conductivity

Note!

This section should be read in conjunction with the calibration section (Chapter 6) and the mainte-

nance section (Chapter 7). The calibration (cell constant) of a sensor does not change unless the

sensor is damaged.

It can also appear to change because of coating of the electrodes, or partial blockage. Because

these changes should be handled as described in the maintenance section, it does not make

sense to regularly recalibrate the ISC450G.

A calibration check, however, is another matter. When the objective is clearly defined as a diagnos-

tic exercise a regular check can bring an extra level of security and confidence to the measure-

ment.

Sensor damage, and/or coatings can be difficult to see and the calibration check can confirm their

presence, by a deviation from the known solution conductivity. The remedial action should be to

clean the sensor, and carefully check for blockage or damage (not simply to recalibrate).

Higher conductivity solutions should be used where possible. The lower the conductivity of the test

solution, the easier it is to contaminate. Carbon dioxide from the air can be quickly absorbed to

cause an error. All containers must be suitably clean, and all materials suitably pure. Outside of a

well-equipped laboratory these conditions are hard to meet.

EXAxt ISC450G is programmed with the following table of conductivity of Potassium Chloride

(KCl) solutions at 25°C. This is used in the Automatic Cell Constant setting calibration feature.

(See chapter 6 on calibration) The table is derived from the Standards laid down in “International

Recommendation No. 56 of the Organisation Internationale de Métrologie Legale”.

Table 11-2. KCl values at 25 °C

mol/l

Advertising