Figure 3-9b. Dimensions

3-6-4. Sensor cable connections using junc-

tion box (BA10) and extension cable

(WF10)

Where a convenient installation is not possible

using the standard cables between sensors and

converter, a junction box and extension cable

may be used. The Yokogawa BA10 junction box

and the WF10 extension cable should be

used. These items are manufactured to a very

high standard and are necessary to ensure that

the specifications of the system can be met.

The total cable length should not exceed 60

metres (e.g. 5 m fixed cable and 55 m exten-

sion cable).

Ø 40 ( 1.57)

wrench opening

20 ( 0.79 )

wrench opening

32 ( 1.42 )

thermistor

ground

secondary

primary

YOKOGAWA

G3/4

47(1.85)

11

12

17

13

15

16

14

L= 5000 ( 200 )

ca 240 (9.45)

40 (1.57)

100 (3.94)

Hole in flange

Ø 27 mm (1.06")

D

d

I

t

d = distance min 25 mm (1")

D = acces port size min 48 mm (1.89")

DIMENSIONS

INSTALLATION INSTRUCTIONS

BULK-HEAD MOUNTING

distance min. 25 mm (1)

access port side min 48 mm (1.89)

UNIT: mm (inch)