This test is done on the “HIF” display of “Factory Mode”.

a. Touch the [Setup] icon.

b. Touch the [Commissioning].

c. Touch the [Advanced setup].

d. Touch the [Factory adjustment].

e. Enter a password.

f. Select the [Factory Mode] in “Key.”

g. Select the [HIF] in “Execute.”

Wind ten turns of wire onto ISC40 sensor. When the resistance of the decade resistance

box 2 to the corresponding value “RANGE1” to “RANGE3” in Table 1 is set, check the

data display and the value must be within the range shown in Table 1.

Table 1

RANGE

Resistance (Ω) of

Decade Resistance Box 2

Data Display (Ω)

RANGE1

150

1.5 ±0.01

RANGE2

1.5k

15 ±0.1

RANGE3

15k

150 ±1

After the above test is completed, wind one turn of wire onto ISC40 sensor.

When the resistance of the decade resistance box 2 to the corresponding value

“RANGE4” to “RANGE6” in Table 2 is set, check the data display and the value must be

within the range shown in Table 2.

Table 2

RANGE

Resistance (Ω) of

Decade Resistance Box 2

Data Display (Ω)

RANGE4

1.5k

1.5k ±0.01k

RANGE5

15k

15k ±0.2k

RANGE6

150k

150k ±12k

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.4 Temperature

Indication Check

Following Section 3.3, select the [PT1000] in “Temperature” of the “HIF” display.

In this state, change the resistance of the decade resistance box 1 and check the data

display. The value on the data display must be within the range shown in Table 3.

Table 3

Temperature (°C)

Resistance (Ω) of

Decade Resistance Box 1

Data Display (°C)

–10

960.9

–10 ±0.3

25

1097.3

25 ±0.3

130

1498.2

130 ±0.3

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.5 Current Output Test

Following Section 3.4, select the [Check] in “mA outputs” of the “HIF” display. “Set value

4.000 mA” appears at the bottom of the display. Select “Next value” in the “Command”

and touch “Enter,” the value on the data display increases in steps of 4 mA. Check the

current outputs 1 and 2 corresponding to the data display, the current output must be

within the range shown in Table 4.