Operation of exaxt isc450g, 1. main display functions, 2. trending graphics – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 24: 3. zoom in on details, Next

Advertising
background image

16

IM 12D06D05-01E

4-1. Main display functions

Go to trend screen

Go to zoom screen
Go to status screen
Go to maintenance screen

Tag

:EXAxt ISC450

3.760

25.0

mS/cm

16.64

+

Figure 4-1. Main Display

A heart “ ” mark is lit on the right-upper corner of the

screen when HART communication is active. An “Χ”

mark is lit when HART communication is abnormal.

Nothing appears when HART communication is not

used.

Note that the “Χ” mark may appear due to output

signal noise or the like even when HART communica-

tion is not used. There is not problem when HART

communication is not used. Continue operation while

ignoring the mark.

4-2. Trending graphics

Pressing the button changes the display

into a graphical mode in which the average

measured value is shown on a time scale. The

“Live” value is also digitally displayed in a text

box. The time scale ( X-axis) and the primary

value scale (Y-axis) are set in the “DISPLAY

SETUP” menu. The full screen displays a trend

of 51 points that represent the average of the

selected time interval. The analyzer samples

the measurement every second. The trending

graphic also shows the maximum and minimum

measured value in that interval.

For example if the time scale is set to 4 hours,

then the trend is shown for 4 hours prior to the

actual measurement. Each point on the trend

line represents the average over 4*60*60/51=

282 measurements (seconds).

Maximum

Minimum

Minimum

Maximum

Average

Live reading

ISC

SC

T

120.0

90.0

60.0

30.0

109.3 µS/cm

450

Figure 4-2. Trend screen

4-3. Zoom in on details

This button gives access to the diagnostic infor-

mation of the analyzer. The following messages

will appear under normal (default) conditions:

- Home key back to
mainscreen.
- One level up.

- Scroll choices (grey
means deactivated).
- Enter selected data
or choice.

First zoom screen
gives you inside
into the parameters
involving current
measurement. All
following zoom
screens give
additional information
about the device and
lead to logbook data.

Next

Next

Next

Next

Figure 4-3. Detail screen

4. OPERATION OF EXAxt ISC450G

Advertising