18

IM 12D06D05-01E

4-3-14. HART Device revision

Sometimes the firmware of a device is

updated in a way that the communication file

(HART DD) need revision too. In this case the

revision level is increased by one. The revision

level of the HART DD must match the revision

level of the Firmware. The revision level is

expressed by the first two characters of the

filename. The following files should be used

when the HART Device revision level is 2.

(

0201.aot, 0201.fms, 0201.imp, 0201.sym)

4-3-15. Logbook

The EXAxt contains several logbooks to store

historical information on events, changed

settings and calibrations. The logbooks have

been categorized to simplify the retrieval of this

information.

Calibration will give information of previous

calibrations. This logbook is useful as one now

can

1) Monitor the sensor performance over time.

2) Monitor the sensor(s) lifetime.

Sensor will give historical information on

parameter settings concerning the sensor(s).

The events logged in this logbook are user

definable. This is done in:

Commissioning >> Configure Logbook >>

Sensor Logbook.

Settings wil give all history information on pa-

rameter settings concerning the analog outputs

(mA1/mA2) and contact (S1…S4). This logbook

is useful to trace back differences in perform-

ance due to changed settings. The events

logged in this logbook are user definable. This

is done in:

Commissioning >> Configure Logbook >>

Settings Logbook – mA and/or Settings

Logbook – contact

mA1/mA2 shows all (dynamic) events concern-

ing the analog outputs

S1/S2/S3/S4 shows all (dynamic) events con-

cerning the contacts.

Each HMI screen can contain up to 5 events.

As each logbook can contain 50 events in total,

one can access previous events by selecting

another page 1 to 10.

4-3-16. Trouble shooting

If you contact the local sales/ service organiza-

tion the serial number and software revision is

necessary information. Without that information

it is impossible to help you. It is also very useful

to report all the information that you find on the

zoom-in display.

4-4. Information function

In this field an information sign , a warning

sign or a fail sign can appear. Pushing

this button, the user gets detailed information

about the status of the sensor or the instrument

if applicable.

See troubleshooting (chapter 8) for further

details.

4-5. Setup-Calibration & commissioning

By pressing the setup key, you get access

to the operating system of the converter based

on menus and submenus.

Browse through the list using the key

till you find the required menu and press the

key to enter this menu.

It is also possible to press on the or

sym-

bol found beside the menu item.

4-6. Secondary- primary value display switch

Pressing on this text block automatically

switches the secondary value to the main

display (Large font size).

25.0