3-2. access to terminal and cable entry, 3-3. ac power, 3-4. dc power – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 18: 3-5. grounding the housing, 3-6. switching on the instrument, Warning, Danger, Power

Advertising
background image

10

IM 12D06D05-01E

3-3-5. Grounding the housing

For the safety of the user and to protect the

instrument against interference, the housing

must always be connected to ground. This has

to be done by a large area conductor. This

cable can be fixed to the rear of the housing or

by using the internal ground connections using

a braided wire cable. See figure 3-8.

The minimum cross sectional area of the pro-

tective grounding wire should be 0.75 mm

2

.

3-3-6. Switching on the instrument

After all connections are made and checked,

the power can be switched on from the power

supply. Make sure the LCD display comes on.

After a brief interval, the display will change to

the measured value. If errors are displayed or

a valid measured value is not shown, consult

the troubleshooting section (Chapter 8) before

calling Yokogawa.

DANGER

Figure 3-7. Input and output connections

63

66

65

61

22

21

12

16

15

-

+

-

+

mA OUTPUTS

62

11

17

TEMP

mA2

SHLD

CONTACT

SENSOR(S)

REFER TO INSTRUCTION MANUAL FOR CONNECTIONS

ISC

mA1

(+HART)

14

COILS

RECEIVE

DRIVE

SHLD

+

-

13

32

31

33

42

41

43

52

51

53

72

71

73

NC

C

NO

NC

C

NO

NC

C

NO

NO

C

NC

S1

S2

S3

CONTACTS

S4

250V / 5A

AC / DC

100VA / 50W

(fail-safe)

2

1

-

+

DC

POWER

12-24 V /10 W

FUSE: 2A/250 VAC/T

2

1

N

L

AC

POWER

100-240 VAC/15 VA/ 50/60Hz

FUSE: 500 mA/250 VAC/T

Fuse replacement should be performed only by

a qualified service person.

See Sec.7. MAINTENANCE, Fuse.

Fuse ratings:

Power supply

Fuse type

12-24VDC, 10W max

2A/250V, Slow

100-240VAC, 15VA max 0.5A/250V, Slow

3-3-2. Access to terminal and cable entry

Terminals 1 and 2 are used for the power sup-

ply. Guide the power cables through the gland

closed to the power supply terminals. The ter-

minals will accept wires of 2.5 mm

2

(14 AWG).

Always use cable finishings if possible.

WARNING

3-3-3. AC power

Connect terminal L to the phase line of the

AC power and terminal N to the zero line. See

figure 3-8 for the power ground.

3-3-4. DC power

Connect terminal 1 to the positive outlet and

terminal 2 to the negative outlet.

The size of conductors should be at least 1.25

mm

2

. The overall cable diameter should be

between 7 & 12 mm.