2. cell constant manual, 3. cell constant automatic, 4. air (zero) calibration – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 46: 5. sample calibration, 6. temperature coefficient calibration, Caution

Advertising
background image

38

IM 12D06D05-01E

6-2. Cell constant manual

The intention of this calibration routine is to fine

tune a sensor for which only the nominal cell

constant is known, or recalibrate a sensor that

has been changed (or damaged) in the course

of operation. Choose 1st or 2nd compensation

to suit the calibration solution used. The solu-

tion should be prepared or purchased, meeting

the highest standards of precision available.

Allow the sensor to reach stable readings for

both temperature and conductivity before ad-

justing to correspond to the calibration solution

value. The setting of a cell constant for a new

(replacement) sensor is also possible in this

routine. This avoids the need for entry into the

commissioning mode, which may have another

authorization (password) level.

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Cell constant (manual)

” to

execute calibration.

6-3. Cell constant automatic

This routine is built around the test method

described in OIML (Organisation Internationale

de Metrologie Legale). International Recom-

mendation No. 56. It allows the direct use of

the solutions prescribed in the test method, au-

tomatically selecting the appropriate tempera-

ture compensation. The look up table is used

to find the appropriate conductivity reading for

the measured temperature. See appendix 3 for

OIML solutions

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Cell constant (automatic)

” to

execute calibration.

6-4. Air (zero) calibration

With the clean dry cell in open air, the reading

should be zero. The Air cal compensates for

excess cable capacitance, and gives a better

accuracy at low readings. This should be done

for all installations during commissioning. After

some time in service a dirty sensor may well

show a high zero offset because of fouling.

Clean the sensor and try again.
• Wait for the sensor to dry (because air cal

must be made with no current flow in the

sensor, that is, the sensor must be dried

while being exposed to air).

• The sensor must be installed in an

environment free of electromagnetic

interference and radio frequency

interference.

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Air calibration

” to execute it.

CAUTION

The temperature compensation should be set

to NaCl when confirming zero during air calibra-

tion.

6-5. Sample calibration

With the sensor in situ, a sample can be taken

for laboratory analysis. Sample calibration

records the time and reading, and holds these

in memory until the analysis has been com-

pleted. The laboratory data can then be entered

regardless of the current process value, without

the need for calculations.

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Sample

” to execute sample

calibration.

Press [

Take Sample

] to record a collected

sample value in memory. Re-enter the Sample

Cal. screen and press [

Start calibration

] to

perform a sample calibration. This updates the

recorded data.

6-6. Temperature coefficient calibration

Simply input the solution conductivity at the

reference temperature (T

R

) after allowing the

sensor to stabilize at elevated temperatures.

The ISC450G will calculate the temperature

coefficient for you. The ideal temperature for

this calibration is the normal process value (T

P

).

This calibration is enabled if the Temperature

Compensation is set to “TC.”

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Temperature coefficient

” to

execute this calibration

Advertising