6. wiring the sensor, 6-1. general precautions, 6-2. connecting the sensor cable to the converter – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 20: 6-3. installation of the sensor

Advertising
background image

12

IM 12D06D05-01E

Figure 3-9a. Dimensions

It is necessary to use screening/shielding on

the output signal cables. Terminal 63 is used to

connect the shielding.

3-6. Wiring the sensor

3-6-1. General precautions

The sensor cable transmits low voltage, high

frequency signals and should be installed

separately from any high voltage, high cur rent

and/or power switching cables. This is to avoid

any unintentional cross talk or other kind of

interference of the conductivity meas ure ment.

3-6-2. Connecting the sensor cable to the

converter

1. To access terminals remove the front cov-

er of the EXAxt ISC450G by releasing the

4 captive screws.

2. Loosen the cable gland and pull the ca ble

in the connection compartment.

3. The sensor leads are numbered and the

leads must be connected to the ter mi nals

with the corresponding number, re fer to

Fig. 3-4: the temperature com pen sa tor

with 11/12

- the drive coil with 15/16

- the receive coil with 13/17

Terminal 14 is for connection of the shield.

4. Screw the cable gland tight to en sure IP66

(NEMA 4X) environmental pro tec tion. An

optional protection hose (flexible con duit)

is available for additional me chan i cal pro-

tec tion of the sensor cable.

3-6-3. Installation of the sensor

The Model ISC40 is a “doughnut” shaped sen-

sor. Preferably, the process will flow through

the hole of the donut with the temperature com-

pensator up-stream. For minimal obstruction of

the flow and for accurate measurement without

the need for calibration of the cell constant, the

process will flow freely around the doughnut,

allowing a minimum distance of 25 mm (1”)

between doughnut and process piping (d).

The sensor is provided with a gasket and

retaining nut. This allows “bulkhead mounting”

in tank wall or standard flange through a hole of

27 mm (1.1”) diameter (A). The insertion is 125

mm under the flange.

Two flats are provided with wrench size 20 mm

(0.8”) to allow easy mounting and alignment of

the sensor. The model identification on one flat

aligns with the “up-stream” position of the sen-

sor.For On-line mounting adapters are avail-

able for standard 2” process connection (Gas

thread, NPT, ANSI-flange, DIN-flange).

For by-pass measurement flow fittings are

available in polyporpylene, polyvinylidene flour-

ide and Stainless Steel.

For measurements in open ducts or vessels an

immersion fitting in CPVC or Stainless Steel is

available.

For easy wiring the sensor must be located

within 5 m (15”) from the converter using the

integral sensor cabling.