5

MENU

STRUCTURE

COMMISSIONING

Menu

Parameter Default

Range

values

min.

max.

Alarm S1

Setpoint

900.0 mS/cm (high)

- inf

+ inf

Alarm S1

Hysteresis

9.000 mS/cm

0.0 µS/cm

+ inf

Alarm S1

Delay Time

0.2 sec.

0.0 sec.

+ inf

Alarm S1

Expire Time

0.0 sec.

0.0 sec.

1800 sec.

Alarm S2

Setpoint

100.0 mS/cm (low)

- inf

+ inf

Alarm S2

Hysteresis

1.000 mS/cm

0.0 µS/cm

+ inf

Alarm S2

Delay Time

0.2 sec.

0.0 sec.

+ inf

Alarm S2

Expire Time

0.0 sec.

0.0 sec.

1800 sec.

S1 (control)

Expire Time

0.0 sec.

0.0 sec.

1800 sec.

P-control S1

Setpoint

500.0 mS/cm

- inf

+ inf

P-control S1

Range

100.0 mS/cm

- inf

+ inf

P-control S1

Manual Reset

0.000%

0%

100%

PID-control S1

I-time

3600 sec.

1.0 sec.

3600 sec.

PID-control S1

D-time

0.0 sec.

0.0 sec.

60 sec.

S2, S3, S4 Similar structure to S1