6-7. Temperature calibration

In order to make the most accurate measure-

ments, it is important to have a precise temper-

ature measurement. This affects the display of

temperature, and the output signal when used.

More important, however, is the temperature

compensation, and calibration accuracy.

The temperature of the sensor system should

be measured independently with a high preci-

sion thermometer. The display should then be

adjusted to agree with the reading (zero offset

calibration only). For best accuracy this should

be done as near to the normal operating tem-

perature as possible.

Press the

key and choose “

Execute: Cali-

bration.

” Press “

Temperature calibration

” to

execute it.

6-8. Operation of hold function during cali-

bration

EXAxt ISC450G has a HOLD function that will

suspend the operation of the control/alarm re-

lays and mA-outputs.

During calibration, the user may choose to

enable HOLD so that the output signals are

frozen to a “

last

” or “

fixed

” value. Some users

will choose to leave the outputs “

live

” to record

the calibration event. This has implications

for pharmaceutical manufacture, for example,

where an independent record of calibrations is

mandatory.

Press to select Execute: “

HOLD

” and then

choose “

Manual Hold ON

” or “

Manual Hold

OFF

”. This allows you to set up manual hold.

To cancel manual hold, press the lit HOLD sec-

tion on the Main display.

Press “

HOLD

” button on mainscreen, to re-

move the HOLD.

The route for HOLD setup is

Commissioning >> Output setup>> Config-

ure Hold