Measure:vmax, Measure:vmin, Measure:vpp – GW Instek GDS-1000A-U Series Programming Manual User Manual
Page 55

COMMAND DETAILS
55
:MEASure:VMAX
Query
Description
Returns the maximum amplitude.
Same as: Measure key → F1~F5 → F3 (Vmax)
Syntax
< Long >
< Short >
:measure:vmax?
:meas:vmax?
Returns
<NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:measure:source 1
:measure:vmax?
Selects Channel 1, and
then measures the
maximum amplitude.
:MEASure:VMIN
Query
Description
Returns the minimum amplitude.
Same as: Measure key → F1~F5 → F3 (Vmin)
Syntax
< Long >
< Short >
:measure:vmin?
:meas:vmin?
Returns
<NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:measure:source 1
:measure:vmin?
Selects Channel 1, and
then measures the
minimum amplitude.
:MEASure:VPP
Query
Description
Returns the peak-to-peak amplitude (difference
between maximum and minimum amplitude)
Same as: Measure key → F1~F5 → F3 (Vpp)