Command overview, Command syntax, Ommand overview – GW Instek GDS-1000A-U Series Programming Manual User Manual
Page 8

GDS-1000A-U Programming Manual
8
C
OMMAND OVERVIEW
The Command overview chapter lists all GDS-
1000A-U commands in functional order. The
command syntax section shows you the basic
syntax rules you have to apply when using
commands.
Command Syntax
Compatible
standard
USB CDC_ACM compatible
SCPI, 1994 (partially compatible)
Command format
trig:del:mod <NR1>LF
1
2
3
4
1: command header
2: single space
3: parameter
4: message terminator
Parameter
Type
Description
Example
<Boolean>
boolean logic
0, 1
<NR1>
Integers
0, 1, 2, 3
<NR2>
decimal numbers 0.1, 3.14, 8.5
<NR3>
floating point
4.5e-1, 8.25e+1
<NRf>
any of NR1, 2, 3 1, 1.5, 4.5e-1
Message
terminator
LF^END
line feed code (hexadecimal 0A)
with END message
LF
line feed code
<dab>^END
last data byte with END message
Note
Commands are non-case sensitive.