Configure gpib interface – GW Instek GDS-3000 Series User Manual User Manual
Page 183

REMOTE CONTROL CONFIG
183
Configure GPIB Interface
To use GPIB, the optional GPIB to USB (GUG-001) adapter must be
used. The GPIB address can be configured for the GUG-001 from
the utility menu. See the GUG-001 user manual for more
information.
Configure GPIB
1. Insert the GUG-001 USB cable into
the rear panel USB device port.
2. Press the Utility key.
Utility
3. Press I/O from the bottom menu.
4. Press USB Device Port from the
side menu.
5. Press GPIB from the side menu.
6. Use the variable knob to set the
GPIB Address from the side menu.
Range
1 ~ 30
GPIB Constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection