Probe Compensation
Panel Operation
1. Connect the probe between the Channel1 input
and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe attenuation to x10 (GDP probes are fixed
at x10).
VOLTS/DIV
VOLTS/DIV
VOLTS/DIV
VOLTS/DIV
POSITION
POSITION
POSITION
TIME/DIV
POSITION
HORIZONTAL
VARIABLE
W
AC
BW
50
W
75
VERTICAL
LEVEL
POSITION
TRIGGER
W
AC
BW
50
W
75
W
AC
BW
50
W
75
W
AC
BW
50
W
75
CH4
CH3
CH2
CH1
EXT TRIG
POWER
2V
Window
Split
Select
Zoom
CH1
CH2
CH3
CH4
R
B1
Measure
Cursor
Display
Help
Test
Save/Recall
Utility
Menu
50 %
Force - Trig
Autoset
Single
Run/Stop
Setup
Auto-Range
Default
Acquire
M
B2
MATH
REF
BUS
Off
Menu
Save
x1
x10
X
1
0
X
1
2. Press the CH1 key to activate CH1.
CH1
3. Set the Coupling to DC from the
bottom menu.
4. Set Impedance to 1MΩ from the
bottom menu.
5. Set the Probe attenuation to
Voltage, 10X.
Page 123
6. Press the Autoset key. The
compensation signal appears on
the display.
Autoset
7. Press the Display key, then
set the display type to
Vector.