SeCtion 2 - installation

2.1

UnPACKinG

Remove the 7310 or 7315 from the packaging and ensure the following items are included:

1. Model 7310 spectrophotometer (731 001), or Model 7315 spectrophotometer (731 501)

2. 24V 65W power supply unit (021 060)

3. Pack of 100 disposable plastic visible wavelength cuvettes (060 084),

or pack of 100 disposable UV plastic cuvettes (060 230)

4. 2 GB USB memory stick (019 146)

5. Jenway 73 series PC software (735 100) and interface cable (013 203)

6. Instruction manual (731 005)

7. Jenway Foreign Manual CD (JENMANCD)

8. Optional accessories (as ordered)

2.2

inStALLAtion

Models 7310 and 7315 are supplied ready to use.

The unit should be placed on a clean flat surface which is free from drafts and vibrations. The units are
designed for operation on 90V to 264V AC input at 47 to 63Hz. Select the correct plug attachment and
attach to the power supply unit as shown below:

Fig 2.2.1 – Power supply unit with various plugs

Connect the power supply unit to the power inlet socket on the rear panel of the instrument and
connect to the mains socket. Turn the power on at the mains and switch the instrument on using the
power switch on the rear of the instrument.

The instrument will perform several power on tests before displaying the main menu: