10.1.2

SAVinG MetHoDS to USB MeMoRY StiCK

In the expanded operating menu press the key adjacent
to the

method icon to open the method selection

menu. Select the method to save to USB memory stick
by pressing the key adjacent to the method location.
Use the keys below the

arrow icons to move page up or

down. Hold the key below the

create method icon for

2 seconds to open the save to UBS memory stick menu.

Either the single method selected or all 48 methods can
be saved to USB memory stick. To save a single method
press the key adjacent to the

single method save icon.

To save all 48 methods press the key adjacent to the

1

- 48 method save icon. Press the key adjacent to the
tick icon to confirm saving the method(s) to USB memory
stick. Press the key adjacent to the

cross icon to cancel

saving the method(s) to USB memory stick and return to
the method selection screen.

Confirmation will be needed to save the methods to USB
memory stick. Press the key adjacent to the

tick icon to

confirm saving the method(s). Press the key adjacent
to the

cross icon to cancel and return to the option of

saving one or all of the methods to USB memory stick.

10.2

oPeninG MetHoDS

Methods can be opened from the internal memory of the instrument or from a USB memory stick.
Before a method can be opened from a USB memory stick it must be saved to the internal memory of
the instrument first.

10.2.1

oPeninG MetHoDS FRoM inteRnAL MeMoRY

In the expanded operating menu press the key adjacent
to the

method icon to open the method selection menu

and use the keys below the

arrow icons to move page

up or down. Select the method to open by pressing the
key adjacent to the method location. Press the key below
the

open method icon.

The instrument will return to the current measurement
mode and the method name will be displayed at the top
of the expanded operating menu.

If a method hasn’t been selected it is still possible to calibrate the instrument and take measurements.
The instrument will operate in a generic method, 0, where all the current operating settings will be
stored. There are separate method 0’s for each measurement mode.

01

02

03

04

TEST

05

06

07

08

1

1-48

1

1-48

1

01

02

03

04

TEST

05

06

07

08

0.000

100.0

400

09:02

ABS

%T

nm

01 TEST

Advertising
This manual is related to the following products: