SeCtion 8 – Quantitation

The quantitation measurement mode enables sample concentrations to be calculated using a standard
curve. In this mode a number of standard solutions covering a range of known concentrations are
measured at a set wavelength. The absorbance or % transmittance of these solutions is plotted to create
a standard curve. Once the standard curve has been created a sample of unknown concentration can be
measured and the concentration calculated using the standard curve.

8.1

MoDe SPeCiFiC PARAMeteRS

The quantitation minimal operating menu enables
calibration to zero absorbance and simple readings to be
taken without changing any measurement parameters or
settings. Pressing the key adjacent to the

JW icon opens

the expanded operating menu.

The quantitation expanded operating menu enables
measurement parameters to be changed. The utility
toolbar on the left hand side of the screen enables access
to printing, print setup options, results, methods and
autologging options. For more details on the different
functions of the utility toolbar refer to section 10.

The

standard curve icon enables the current standard curve and curve statistics to be viewed or a new

standard curve to be created. The

quantitation table icon enables the quantitation standards to be

viewed.

Fig 8.1 - Expanded Operating Menu – Post Measurement

Minimal Operating Menu

Expanded Operating Menu

-0.000

0.000

400

09:02

ppm

ABS

nm

-0.000

0.000

400

09:02

ppm

ABS

nm

0.986

0.000

400

09:02

ppm

ABS

nm

Print/print settings

Standard curve

Quantitation table

Increase wavelenth

Decrease wavelength

Calibrate

to zero

Measure

sample

Results selection menu

Method selection menu

Autolog menu