Jenway 73 Series PC software User Manual

Page 23

Advertising
background image

23

4.1.7

Auto Logging

Auto

logging

allows

repeat

measurements of the same sample

to be performed with a set time

interval between each measurement.

Use the arrow icons to increase or

decrease the time interval between

sample measurements and use the

arrow icons to select the number of

readings required. If auto logging is

not required the values do not need

to be adjusted in this screen.

4.2

CALIBRATION AND SAMPLE MEASUREMENT

In the concentration measurement mode calibrations against a standard or a factor can be performed

following a zero calibration. If the factor is not known calibration against a known standard is

performed in order to calculate the factor. However if the factor is known there is no need to calibrate

using a standard. The calibrations must be performed at the same wavelength at which the sample

will be measured.

4.2.1

Calibration and sample measurement using a factor

Once the settings have been entered the zero calibration should be performed.

Insert the cuvette containing the blank sample into the sample chamber and close the instrument lid.

Click on the Zero button; this sets the instrument to zero absorbance.

Once a calibration is complete the

Read and Auto Log buttons are

activated. If the settings menu is re-

entered following the zero calibration

it is not possible to change the

wavelength. If the wavelength needs

to be adjusted another tab should be

opened.

If the settings menu is re-entered following the zero calibration it is possible to change the