Jenway 73 Series PC software User Manual

Page 48

Advertising
background image

48

To continue click Yes, to cancel click No or Cancel.

If the wavelength is changed then the zero calibration must be performed again before a sample can

be measured. The standard samples should also be re-measured at the new wavelength.

Alternatively a new quantitation tab can be opened and the required settings can be entered and a

new quantitation curve can be created.

6.4 SAMPLE MEASUREMENT

Insert the cuvette containing

the unknown sample into the

sample chamber and close

the instrument lid.

Click on the Read button; the

instrument

performs

the

measurement and result is

displayed on the screen.

If the same sample needs to

be measured more than

once click on the Auto Log

button.

The instrument will perform

the required number of

readings with the specified

time interval between each

measurement as set in the

Settings screen.

The results will be displayed on the screen. The batch name entered in the settings screen will have

01, 02 etc after the name, to signify the order in which the results were generated.

Auto logging can be stopped before all the measurements have been completed by clicking on the

Stop button. To recommence auto logging click on the Auto Log button again.