Jenway 73 Series PC software User Manual

Page 47

Advertising
background image

47

Once the standards have

been measured click on

Show Curve to view the

quantitation curve. The curve

fit can be adjusted in this

screen. Click on Close

window to return to the

create curve screen.

Click on OK to save and

return to the measurement

mode.

6.3 CALIBRATION

Once the settings have been entered and if required a new quantitation curve has been created, the

zero calibration should be performed.

Insert the cuvette containing the blank sample into the sample chamber and close the instrument lid.

Click on the Zero button; this sets the instrument to zero absorbance and 100% transmittance.

Once a calibration is complete the

Read and Auto Log buttons are

activated.

If the settings menu is re-entered following the zero calibration it is possible to change the

wavelength. However a warning will be displayed to warn that critical settings are being changed after

the zero calibration has been performed and a quantitation curve has been created.