Click on the Standard button.

The instrument will take a

reading and calibrate to the

standard

concentration.

Once

the

calibration

is

complete the Read and Auto

Log buttons are active and

the

sample

can

be

measured.

If the settings menu is re-entered following the standard calibration it is possible to change the

wavelength. However a warning will be displayed to warn that critical settings are being changed after

the zero calibration has been performed.

To continue click Yes, to cancel click No or Cancel.

If the wavelength is changed then the zero and standard calibration must be performed again at the

new wavelength, before a sample can be measured. Alternatively a new concentration tab can be

opened and the required settings can be entered.