Jenway 73 Series PC software User Manual

Page 57

Advertising
background image

57

7.1.9

Auto Logging

Auto

logging

allows

repeat

measurements of the same sample

to be performed with a set time

interval

between

each

measurement. Use the arrow

icons to increase or decrease the

time interval between sample

measurements and use the arrow

icons to select the number of

readings required. If auto logging is

not required the values do not

need to be adjusted in this screen.

7.2

CALIBRATION

In the time scan kinetics mode the calibration is a zero calibration performed at the selected

wavelength that the sample is going to be measured at.

Insert a cuvette containing the blank solution

into the sample chamber and close the instrument lid.

Click on the Zero button to perform

the zero calibration. Once the

calibration is completed the Read

and Auto Log buttons become active

and the sample can be measured.

The calibration status will show that

the instrument is calibrated.

If the settings menu is re-entered following the zero calibration it is possible to change the

wavelength. However a warning will be displayed to warn that critical settings are being changed after

the zero calibration has been performed.