Installation for flow measurement, 5 connect the sensor cable, Optisonic 6300 p – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 30: Signal strength qualification
Advertising

3
INSTALLATION FOR FLOW MEASUREMENT
30
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
3.4.5 Connect the sensor cable
Figure 3-16: Connecting signal cables
1 Connector for "UP" transducer (blue)
2 Connector for "DOWN" transducer (green)
1
2
3
4
5
6
7
8
9
+/-
0
EXP
ABC
DEF
GHI
JKL
MNO
PQRS
TUV
WXYZ
A/a
-
I/O
DISPLAY
LOGGER
MENU
1
1
2
2
1
2
INFORMATION!
After mounting the transducers in the proper position, it takes up to 60 seconds to find a signal.
INFORMATION!
The signal strength is given and can be qualified as shown:
Signal strength
Qualification
> 75%
High signal
50...75%
Fairly high signal
10...50%
Low signal
< 10%
Bad or no signal
Check settings in menu, change transducer distance
until there is at least a low signal.