Electrical connections – KROHNE OPTISONIC 6300 P-6400 EN User Manual
Page 50

5
ELECTRICAL CONNECTIONS
50
OPTISONIC 6300 P
www.krohne.com
03/2013 - 4000972603 - MA OPTISONIC 6300 P R03 en
• If the following maximum load resistance R
L, max
is exceeded, the load resistance R
L
must be
reduced accordingly by parallel connection of R:
f ≤ 100 Hz: R
L, max
= 47 kΩ
f ≤ 1 kHz: R
L, max
= 10 kΩ
f ≤ 10 kHz: R
L, max
= 1 kΩ
• The minimum load resistance R
L, min
is calculated as follows:
R
L, min
= (U
ext
- U
0
) / I
max
Status output
• U
ext
≤ 32 VDC
• I ≤ 100 mA
• R
L, max
= 47 kΩ
R
L, min
= (U
ext
- U
0
) / I
max
• open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I ≤ 10 mA
U
0, max
= 2 V at I ≤ 100 mA
• The output is open when the device is de-energized.
Figure 5-8: Pulse frequency output active P
a
C+
C-
D
D-
INFORMATION!
•
Any connection polarity.
•
Galvanically isolated.
Figure 5-9: Status output / limit switch passive S
p