Conductivity and tds calibration, 1 automatic or manual calibration, Millivolt (mv) offset adjustment – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 10: Automatic or manual calibration

Advertising
background image

6

4.2 Millivolt (mV) Offset Adjustment
Oxidization Reduction Potential (ORP or Redox) is not a precise

measurement, but is useful as a relative indicator. As such, mV offset

adjustment is not meant to enhance accuracy, but rather to make

readings comparable to a reference. Commercial ORP solutions are

often used as a check standard in which a meter/electrode system are

verified to be close to a given value, instead of being used as a

calibration standard in which adjustments are made in an attempt to

match the ORP value.

1) Connect an ORP electrode and press

as needed to select mV

(or R.mV).

2) Dip the ORP electrode into a solution with a known mV value

(i.e. Zobel, Light’s, quinhydrone, or iodidetriiodide) and stir.

3) Press

when the reading is stable. The primary display shows

the relative millivolt value (R.mV) while the secondary display shows

the factory default mV value.

4) Adjust the R.mV value using

or

. Press

to accept or

to cancel. The meter allows an adjustable maximum value of ±

150 mV from the factory default mV value.

Note: When an offset has been stored successfully, R.mV replaces

mV.

5. CONDUCTIVITY AND TDS CALIBRATION

5.1 Automatic or Manual Calibration
The PC 700 is capable of automatic or manual calibration for

conductivity, and manual calibration for TDS.
In the automatic calibration mode, the meter will automatically select

one of (4) conductivity calibration standard values depending on the

range and normalization temperature being used (

TABLE 1).

Advertising