7 3.3 atc (auto temp compensation)—con & tds only, 8 3.4 tds (tds factor)—con & tds only, 3 atc (auto temp compensation)—con & tds only – Thermo Fisher Scientific Eutech PC 700 User Manual
Page 18: 4 tds (tds factor)—con & tds only, 5 t.co (temperature coefficient)—con & tds only

14
9.7 3.3 AtC (Auto Temp Compensation)—Con & TDS only
Press
.
Press
or
to select “Yes” or “No”.
Press
to confirm.
9.8 3.4 tdS (TDS factor)—Con & TDS only
Press
.
Press
or
to select the desired TDS factor (.40 to
1.00).
Press
to confirm.
9.9 3.5 t.CO (Temperature Coefficient)—Con & TDS only
The temperature coefficient is the amount of change in conductivity per
degree temperature (% per ºC). The PC 700 is factory set to a
temperature coefficient of 2.1 % per ºC. For most applications this will
provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature
will be measured by the electrode and displayed in measurement
mode—without compensation.
– Calculating Temperature Coefficients.
Press
.
Press
or
to select the desired value (0.00 to 10.0).
Press
to confirm.