Conductivity electrode, Ph and mv calibration, Ph calibration – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 8

Advertising
background image

4

3. CONDUCTIVITY ELECTRODE

The PC 700 includes an electrode with a nominal cell constant of k

= 1.0, built-in temperature sensor, and 1 meter cable. The Ultem

body housing has good chemical resistant properties. The

electrode design offers fast temperature response and reduces air

entrapment, ensuring accurate, repeatable, and stable readings.
The wetted materials of the probe include:

Polyetherimide (Ultem) – protective probe guard

Polybutylterphalate (Valox) – sensor housing

Stainless Steel (SS 304) – 2 bands

The protective probe guard can be removed temporarily for

cleaning however it must be re-attached during measurement and

calibration.

Erroneous results will occur while the probe guard

is removed
Always immerse the probe beyond upper steel band for best

results. Use the fill line on the outside of the probe guard for

reference.

.

4. PH AND MV CALIBRATION

4.1 pH Calibration
For best results, periodic calibration with known accurate standards is

recommended. Calibrate with standards that bracket your intended

measuring range while including a neutral standard (pH 7.00 or 6.86).

For example, if you expect to measure samples from pH 6.2 to pH 9.5,

calibration with 4.01, 7.00, and 10.01 will work well.

The PC 700 meter can be calibrated with up to 5 buffers. The

non-volatile memory retains all calibration values upon meter shut

down.

Advertising