be available to view. If no calibration is stored, the offset will be 0.0
mV and the slope is 100%.
9.3 3.0 ConF (Configuration)
Press
to access set-up programs 3.1 thru 3.9.
9.4 3.1 rdY (Ready / Stability Indicator)
Press
.
Press
or
to choose READY “On”, READY “OFF”, or
Auto HOLd.
Press
to confirm.
9.5 3.2 ºC ºF (Celsius or Fahrenheit)
Press
.
Press
or
to select ºC or ºF.
Press
to confirm.
9.6 3.3 buFF (pH Buffers & Calibration Points)—pH only
Press
.
Press
or
to select “USA” or “NIST” buffer group
(pH mode only).
Press
or
to select number of Calibration Points
(pH and Ion only).
Press
to confirm.