13 3.8 cell (nominal cell constant)—con & tds only, 8 spc (single point calibration)—con & tds only, 8 cell (nominal cell constant)—con & tds only – Thermo Fisher Scientific Eutech PC 700 User Manual
Page 19

15
9.10 3.6 t.nr (Normalization Temperature in ºC)—Con & TDS only
When Automatic Temperature Compensation is used, measurements
are adjusted by the temperature coefficient to the normalization
temperature. The default value is 25 ºC.
Press
.
Press
or
to select the desired value (15.0 to 30.0).
Press
to confirm.
9.11 3.7 ACAL (Auto Conductivity Calibration)—Con & TDS only
for more details on Automatic Conductivity Calibration.
Press
.
Press
or
to select “Yes” (Automatic) or “No”
(Manual).
Press
to confirm.
9.12 3.8 SPC (Single Point Calibration)—Con & TDS only
for details on Single Point & Multi Point Calibration.
Press
.
Press
or
to select “Yes” (Single) or “No” (Multi).
Press
to confirm.
9.13 3.8 CELL (Nominal Cell Constant)—Con & TDS only
The PC 700 includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved
performance in extreme sample ranges. Use this setup function to
change the cell constant if necessary. Meter default is 1.0 to match the
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).