13 3.8 cell (nominal cell constant)—con & tds only, 8 spc (single point calibration)—con & tds only, 8 cell (nominal cell constant)—con & tds only – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 19

Advertising
background image

15

9.10 3.6 t.nr (Normalization Temperature in ºC)—Con & TDS only
When Automatic Temperature Compensation is used, measurements

are adjusted by the temperature coefficient to the normalization

temperature. The default value is 25 ºC.

Press

.

Press

or

to select the desired value (15.0 to 30.0).

Press

to confirm.

9.11 3.7 ACAL (Auto Conductivity Calibration)—Con & TDS only

See Section 5.1

for more details on Automatic Conductivity Calibration.

Press

.

Press

or

to select “Yes” (Automatic) or “No”

(Manual).

Press

to confirm.

9.12 3.8 SPC (Single Point Calibration)—Con & TDS only

See Section 5.2

for details on Single Point & Multi Point Calibration.

Press

.

Press

or

to select “Yes” (Single) or “No” (Multi).

Press

to confirm.

9.13 3.8 CELL (Nominal Cell Constant)—Con & TDS only
The PC 700 includes a probe with a nominal cell constant (k) of 1.0.

Use probes with k = 0.1 and 10 (sold separately) for improved

performance in extreme sample ranges. Use this setup function to

change the cell constant if necessary. Meter default is 1.0 to match the

included probe.

k = 0.1 ideal for low measurements <20 µS (<10 ppm).