14. SPECIFICATIONS
pH Range
-2.00 to 16.00 pH
Resolution
0.01 pH
Accuracy
±
0.01 pH
Calibration Points
Up to 5 points with Auto-buffer recognition
Buffer Options
USA : pH 1.68, 4.01, 7.00, 10.01,12.45
NIST: pH 1.68, 4.01, 6.86, 9.18 ,12.45
Slope Display
Yes (with offset)
mV Range
±
2000 mV
Resolution
0.1 mV (± 199.9 mV), 1 mV beyond ± 200 mV
Accuracy
±
0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200 mV
Offset Adjustment
Up to ±150 mV
Conductivity Range
..µS to 200.0 mS
Conductivity
Resolution
0.01 / 0.1 / 1 μS ; 0.01 / 0.1 mS
TDS Range
(0.5 TDS factor)
.. to 100.0 ppt @ 0.5 fact (200.0 @ 1 factor)
Resolution
0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt
Accuracy
±1% Full Scale
Temperature
Range (Meter)
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Compensation
Type & Range
Automatic with supplied cell or Manual.
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Resolution
0.1 ºC / ºF
Accuracy
± 0.5ºC / ± 0.9ºF
Calibration
Offset in 0.1 º increments; Offset range: ± 5.0 ºC / 9.0 ºF
Coefficient
0.0 to 10.00 per ºC