Thermo Fisher Scientific Eutech PC 700 User Manual

Page 25

Advertising
background image

21

PROBLEM

CAUSE

SOLUTION

Calibration error.

Buffer value does not match

displayed value or electrode is

disconnected or failing.

Use fresh buffer solutions.

Check electrode connection.

Clean & recondition electrode.

Replace electrode.

Unstable pH

reading

Broken or worn electrode

Replace electrode

External ‘noises’ or induction

(electrical ‘noise’ caused by a

nearby motor)

Remove or switch off interfering device

Slow response

Dirty electrode.

Clean & rehydrate the electrode

CAL

during pH CAL

mode

Calibration was attempted

using only 1-point that was not

pH 7 (USA) or pH 6.86 (NIST).

Recalibrate using 2 or more pH points

or use pH 7 (USA)

or use pH 6.86 (NIST).