5 manual conductivity & tds calibration procedure, 6 temperature calibration, Manual conductivity & tds calibration procedure – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 13: Temperature calibration

Advertising
background image

9

4) When the

READY indicator appears, press

to accept. The

primary reading will flash briefly before returning to measurement

mode upon successful calibration.


5.5 Manual Conductivity & TDS Calibration Procedure

1) Press

as needed to select conductivity (μS or mS) or TDS

(ppm or ppt) calibration.

2) Dip the electrode into the calibration standard and press

.

Provide stirring for best results.

3) The primary display will show the current reading, while the

secondary display will be the factory default value. Adjust the

conductivity or TDS reading using

or

. Press

to

accept. The primary reading will flash briefly before returning to

measurement mode upon successful calibration.


5.6 Temperature Calibration
The thermistor sensor used for automatic temperature compensation

and measurement is accurate and stable, so frequent calibration isn’t

required. Temperature calibration is recommended upon probe

replacement, whenever the temperature reading is suspect, or if

matching against a certified thermometer is desired.

1) Place the probe into a solution with a known accurate temperature

such as a constant temperature bath.

2)

Press

followed by

. The primary display shows the

measured temperature while the secondary display shows the

factory default temperature.

4)

Adjust the temperature using

or

. Press

to accept or

to cancel. The meter allows an adjustable maximum value of ±

5

º

C (or ± 9

º

F) from the factory default temperature.

Advertising