Thermo Fisher Scientific Eutech PC 700 User Manual

Page 9

Advertising
background image

5

The following calibration standards are automatically recognized;

See Section 9.6 to change the buffer group

1) Press

as needed to select pH.

2) Dip the pH and ATC electrodes into pH buffer and press

. The

secondary display will lock on the appropriate buffer value. Provide

stirring for best results. When the

READY indicator appears, press

to accept. The primary reading will flash briefly before the

secondary display begins scrolling the remaining available buffers.

3) Rinse the pH and ATC electrodes then dip into the next pH buffer.

The secondary display will lock on the appropriate buffer value.

When the

READY indicator appears, press

to accept. The

primary reading will flash briefly then display the percent efficiency

(slope) before the secondary display begins scrolling the remaining

available buffers.

4) To calibrate another buffer repeat step 3) or press

to return to

the measurement mode.

Note: The meter will automatically return

to measurement mode upon successful completion of the number of

specified calibration points. To specify a different number of pH

calibration points see Section 9.6.

TIPS: A single point (offset) calibration is only allowed with pH 7.00 or

pH 6.86 buffers.

When the first calibration value is accepted during a new calibration,

all prior calibration values are erased.

Press

at any time to abort calibration and return to measurement

mode.

USA buffer group

1.68, 4.01, 7.00, 10.01, 12.45

NIST buffer group

1.68, 4.01, 6.86, 9.18, 12.45